Modus | DC, VENA et EXAMINO |
Operatio Temperature | 10-40℃ (tantum solum tectis) |
Salus et EMC | Iuxta Ec- cipia;Emissio Group I Classis A;cMETus probatus |
Sensoriis | Ad III fluxus cellulae |
Max Current Compensation (Autozero | - 25 nA-25 mA in DC et PULSUS modus ab occasu range dependens |
Clibanus | +7°C supra ambiens ad 60°C, accuratissime 0.5°C, firmitatis 0.1°C;accommodat columnam et cellam fluunt (s) |
Offset | +50% ad - 50% of max.output voltage, 5% steps |
Analog Output (DAC) | -1 ad I V magnitudine (per XVI frenum D / A converter) |
Analog Output (I/E) | -2.5 ad +2.5 V magnitudine |
DC modus | |
Range | 10 pA-200 uA in 1. 2. 5 incrementa |
Filtrum (ADF) | 10-0.001 Hz in 1.2.V incrementa INCOCTUS et intempesta: nam unprocessed notitia |
Potentia (EC) | -2.50V ad +2.50V cum 10 mV incrementis |
Data Rate | 1- 100 Hz in 1, 2, 5, incrementis colum dependens |
sonitus | <2pA cum cellula dummy (onere 300 MΩ /470 pF) in 1 nA range, eliquare, Ec+800mV et temperaturam 35'C |
PULSATIO Modus | |
Range | 10 nA - 200 µA in 1, 2, 5 incrementis Off: pro unprocessed notitia |
Filtrum (ADF) | 0.5 - 0.001 Hz in 1, 2, 5 incrementis |
Potentia (Ec) | -2.50 V ad + 2.50 V cum 10 mV incrementis |
Data Rate | 1/(venae durationis) Hz |
EXAMINO Modus | |
Range | 10 nA - 200 µA in 1, 2, 5 incrementis |
Potentia (Ec) | -2.50 V ad + 2.50V cum 10 mV incrementis |
Data Rate | 1 Hz Scan rate 1 - 100 mV/s in 1, 2, 5 incrementa |
alii | |
Cycle | Dimidium, plenum, continuum |
Dimensiones | 43(D)*22(W)*44(H)cm* |
Potentia requisita | 100-240 VAC, 50/60 Hz, 260 VA, auto- sentientia |
Pondus | Max 14. 4 kg sine cellula et columna influunt |